Model 245 Pressure Transmitter DigiComp
0-5 VDC, High Accuracy, 3 thru 5000 psi
Features
High Accuracy
Many Options available
All Stainless Steel construction
Shunt Calibration circuit option
Low Pressure Applications
- Description
- PERFORMANCE
- ELECTRICAL
- Attachment
Description
Viatran’s Model 245 DigiComp pressure sensor is among our most accurate pressure transmitters for industrial test and research applications. The 245 measures gage and absolute pressure from 3 PSI to 5,000 PSI (0.2 to 345 bar) with
standard accuracy of ≤±0.06% FSO.
Model 245 has external adjustments which allow the user to fine tune the zero reading by use of a potentiometer. It also has adjustment for span and an installed calibration circuit allowing for field calibration. 245 can be customized for outputs anywhere between -10 to +10Vdc, with limitless electrical and process connection options.
Data sheet 245-345 Digicomp
Full Scale Pressure Ranges
0-3 thru 5000 PSIG (PSIS FOR >500 PSI) (0.2 thru 345 bar)
0-15 thru 0-5000 PSIA (0-1.0 thru 0-345 bar)
0-15 PSIV (0-1.03 BarV)
Nonlinearity (Best Fit Straight Line)
≤±0.06% FSO
Hysteresis & Repeatability
≤±0.06% FSO
Full Scale Output (FSO)
5 Vdc at 70°F (21°C) ≤±0.10% FSO standard
Resolution
Infinite
Long Term Stability
≤±0.3% FSO per 6 months
Compensated Temperature Range
32°F to 170°F (0° to 77° C)
Process Media Temperature Range
-40°F to 250°F (-40°C to 121°C)
Ambient Operating Temperature Range
-40°F to 185°F (-40°C to 85°C)
Storage Temperature Limits
-40°F to 250°F (-40°C to 121°C)
Temperature Effect on Zero
≤±1.0% FSO per 100°F
Temperature Effect on Span
≤±1.0% FSO per 100°F
Supply Voltage
8 to 30 Vdc at 3.5 mA nominal, 5 mA max
(12 to 30 Vdc for 10 Vdc output option)
Power Supply Regulation
≤±0.01% FSO per volt change over the supply voltage range
Output Signal
0 – 5 Vdc standard (see options)
Load Resistance
100K Ohms min
Zero Adjustment
±5% FSO min
Span Adjustment
±5% FSO min
Calibration Signal
80% of FSPR, by shorting calibration pins
Calibration Signal Accuracy
≤±0.1% of the stated value
Circuit Protection
Input polarity may be reversed. Output may be short-circuited indefinitelyOver voltage protection to 1000 volts according to EN61000-4-5
Insulation Resistance
≥ 200 MegOhms to case ground
Response Time
<2 mSec to reach 90% of full scale
RFI / EMI Suppression
CE EMC compliant per IEC EN 61326. CE marked
Electrical Connection
Bendix PT02E-10-6P, mates PT06E-10-6S (SR)
Pin Outs
Pin A + Power
Pin B – Power
Pin C + Signal
Pin D – Signal
Pin E Calibrate
Pin F Calibrate