Model 345 Pressure Transmitter DigiComp

4-20 mA output, High Accuracy, 3 to 5000 psi

Features

 High Accuracy
 Many Options available
 All Stainless Steel construction
 ShuntCal

 

   

Description

Viatran’s Model 345 DigiComp pressure sensor is among our most accurate pressure transmitters for industrial test and research applications. The 345 measures gage and absolute pressure from 3 PSI to 5,000 PSI (0.2 to 345 bar) with
standard accuracy of ≤±0.06% FSO.

The Model 345 has external adjustments which allow the user to fine tune the zero reading by use of a potentiometer. It also has adjustment for span and an installed calibration circuit allowing for field calibration.  In addition to a high accuracy option between temperature range of -40°F to 185°F, 345 has a limitless array of electrical and process connection options to meet any pressure measurement requirement.

 

Data sheet  245-345 Digicomp

Full Scale Pressure Ranges
0-3 thru 5000 PSIG (PSIS FOR >500 PSI) (0.2 thru 345 bar)
0-15 thru 0-5000 PSIA (0-1.0 thru 0-345 bar)
0-15 PSIV (0-1.03 BarV)
Nonlinearity (Best Fit Straight Line)
≤±0.06% FSO
Hysteresis & Repeatability
≤±0.06% FSO
Full Scale Output (FSO)
16 mA at 70°F (21°C) ≤±0.10% FSO
Resolution
Infinite
Long Term Stability
≤±0.3% FSO per 6 months
Compensated Temperature Range
32°F to 170°F (0° to 77° C)
Process Media Temperature Range
-40°F to 250°F (-40°C to 121°C)
Ambient Operating Temperature Range
-40°F to 185°F (-40°C to 85°C)
Storage Temperature Limits
-40°F to 250°F (-40°C to 121°C)
Temperature Effect on Zero
≤±1.0% FSO per 100°F
Temperature Effect on Span
≤±1.0% FSO per 100°F
Supply Voltage
8 to 30 Vdc
Power Supply Regulation
≤±0.01% FSO per volt change over the supply voltage range
Output Signal
4 to 20 mA
Load Resistance
1050 Ohms max at 30 Vdc
Zero Adjustment
±5% FSO min
Span Adjustment
±5% FSO min
Calibration Signal
80% of FSPR, by shorting calibration pins
Calibration Signal Accuracy
≤±0.1% of the stated value
Circuit Protection
Input polarity may be reversed. Output may be short-circuited indefinitelyOver voltage protection to 1000 volts according to EN61000-4-5
Insulation Resistance
≥ 200 MegOhms to case ground
Response Time
<2 mSec to reach 90% of full scale
RFI / EMI Suppression
CE EMC compliant per IEC EN 61326. CE marked
Electrical Connection
Bendix PT02E-10-6P, mates PT06E-10-6S (SR)
Pin Outs
Pin A               + Power / Signal
Pin B                – Power / Signal
Pin C                Calibrate
Pin D               Calibrate
Pin E               No connection
Pin F               No connection